Specifications
Power requirements: 100-250V, 50-60Hz, 2.2A
Measurement units: Force (grams), Travel (thousandths)
Throat dimensions: Standard height = 2.5” (63.5mm)
Displacement (travel)
Range: Displacement of 15mm (custom ranges available)
Resolution: .0001” (.0025mm)
Absolute accuracy: +/- .0015” (+/-.05mm) max
Repeatability: +/- .0005” (+/- .03mm) max
Force
Range: 0-850g
Resolution: 1g (.04oz)
Absolute accuracy: +/- 5g
Repeatability: +/- 2.5g
Tested Parameters
Force Measurements
Fmax = trip force
Fmin = minimum force after Fmax
Fc = force at contact
Fb = force at break
Frmin = minimum force on return
Frmax = maximum force on return
Tactile Slope = Fmax (avg) - Fmin (avg) / Tfmax (avg) - Tfmin (avg)
Travel Measurements
Tfmax = travel to Fmax
Tfmin = travel to Fmin
Tc = travel to contact
Tb = travel to break
Trmin = travel to Frmin
Trmax = Travel to Frmax
Resistance measurements
Tf = final resistance (closed)